Section 5, Maintenance and Service Tests
Allow the current to stabilize before reading the DMM.
Verify that the SUPPLY CURRENT reading on the DMM
is
75 mA or less (1 mV = 1 mA). Ifusing the PVT Form, record the
results.
Turn the power switch off, allow the pump to power down
completely, and remove the battery substitution block. Disconnect
the torque brake and remove the cartridge housing.
AIM/AIM Plus Technical Service Manual
5-29
430-600035-002 (Rev. 11/99)