Figure 5-3. Optics Test Block Placement - Abbott aim Technical & Service Manual

Table of Contents

Advertisement

Section 5, Maintenance and Service Tests
5.3.2.14
Optics Check
I
OPTICS
CHECK
AIR
###
OCCL
###
PRESS
NO
TO
EXIT
Verify the AIR reading is 231 - 255
Verify the OCCL reading is 231 - 255.
Press [NO] to exit.
AIR/OCCL
TEST
\YES'=
PASS/NEXT
'NO'
= FAIL/QUIT
or
'BACKUP'
If using the PVT Form, record the results.
Press [YES] to indicate PASS and go to the next test, [NO] to exit
the FINAL ATP, or press [BACK-UP] to repeat the test.
5.3.2.15
Optics Check li
The second portion of the Optics Check requires an optics test block. If an optics test block is not
available, press [NO] to exit, then [YES] to go to the next test.
&
Test Setup (refer to Figure 5-3. Optics Test Block Placement):
-
Check the optics test block to ensure that the optical surfaces are clean and not
scratched.
-
Position the pump with the cartridge channel facing up.
-
Insert the optics test block into the optics carrier. The arrows must be pointing toward
the bottom of the pump.
Optics block
430-600035-002 (Rev. 11/99)
Optics carrier
mounted on
motor frame
Figure 5-3. Optics Test Block Placement
5-14
AIM/AIM Plus Technical Service Manual