Test Setup
1.
AIM/AIM Plus Technical Service Manual
Section 5, Maintenance and Service Tests
Press [5] and [ENTER] to select a 5 minutes bolus lockout.
Press [3] to set no dose limit.
Press [1] [0] [0] [ENTER] to select a 100 mL container size.
Press [1] to select high air sensitivity.
Press [YES].
Press and hold the down arrow key to scroll through the program
parameters until the review is complete.
At the last screen, press [ENTER] to exit the program review.
No response required. Pump enters the stop mode.
Install a primed cartridge set into the pump. If necessary, refer to the pump's System Operating
Manual for detailed instructions regarding cartridge installation and priming.
Position the proximal end of the tubing in the beaker of water and the distal end in a container to
capture the liquid delivered.
Occlude the distal end of the set by clamping off the tubing.
5-25
430-600035-002 (Rev. 11/99)