5.3.2.20
Encoder Test - AIM Plus Only
Perform either Dry or Wet Test.
Section 5, Maintenance and Service Tests
The AIM Plus Encoder Test requires a turns counter and cartridge housing or primed PROVIDER set
and 20 mL graduated cylinder.
A Dry Test Setup (with turns counter):
-
Install the cartridge housing on the motor frame to defeat the occlusion and check
cartridge alarms.
-
Connect the turns counter to the pump's motor shaft extension.
-
Press RST on the turns counter to reset the display to 0.0.
A Wet Test Setup (with primed PROVIDER set):
-
Install a primed set into the motor frame.
-
Place the proximal end of the set into a water source.
-
Place the distal end of the primed set into the mouth of a 20 mL graduated cylinder.
ENCODER
TEST
0 .0mL
INFUSED
PRESS.
START
PRESS
NO
TO
EXIT
RATE: ###.#
mL/hr
20.0mL
INFUSED
TIME:
179
sec
PRESS
NO
TO
EXIT
ENCODER
TEST
"YES'=
PASS/NEXT
'NO'
=
FAIL/QUIT
or
'BACKUP'
AIM/AIM Plus Technical Service Manual
Press [START] to begin delivery. The volume infused will begin to
accrue.
Pump will run until 20 mLs are delivered.
Verify that the RATE reading is 396.0 - 404.0 mL/hr.
Verify that the TIME reading is 178 - 182 secretary.
For Dry Test:
Verify that the turns counter volume reading is 19.3 - 20.2 mL.
For Wet Test:
Verify that the volume delivered to cylinder is 19.8 - 20.2 mL.
Press [NO] to exit.
If using the PVT Form, record the results.
©
Press [YES] to indicate PASS and go to the next test, [NO] to exit
the FINAL ATP, or press [BACK-UP} to repeat the test.
Press [YES] to exit the FINAL ATP.
5-19
430-600035-002 (Rev. 11/99)