Troubleshooting Tables; Alert Messages - Abbott aim Technical & Service Manual

Table of Contents

Advertisement

Section 6, Troubleshooting
3.
Cycle the power off and on and verify the date is February 29 on the self-test display.
Note:
The time cannot be set when the pump displays February 29. If the correct time takes
precedence over the correct date, refer to the following procedure.
-
To set the correct time when the infused date displayed is February 29, proceed as follows:
1.
Set the date to February 28 or March 1.
2.
Set the correct time.
3.
On March 1, reset the correct time and date.
6.3
Troubleshooting Tables
The following tables are provided to assist in troubleshooting problems with the pump. If problems
require component or assembly replacement, refer to Section 7, Replaceable Parts and Repairs. Also
refer to the PVT in Section 5, Maintenance and Service Tests to aid in troubleshooting the pump.
6.3.1
Alert Messages
Following are descriptions and remedies of alert messages. Press the [HELP] key for information
about an alert message. As necessary, use the arrow keys to scroll through the message. Press the
[BACK-UP] key to exit the help screen.
430-600035-002 (Rev. 11/99)
6-2
AIM/AIM Plus Technical Service Manual