Section 6, Troubleshooting
Table 6-3. AIM/AIM Plus Alarm Codes 008 - 143
Code
Cause
Troubleshooting (repair as required)
135-SP] CHECKSUM
SPI packet
checksum error.
136-SPI ASCII
SPI packet not
ASCH.
137-SPI TRANSFER
SPI packet transfer
error.
138-SPI WAITING
SPI packet wait
state error.
139-SPI PROCESS
SPI packet
processing error.
140-SPI ENDTRANS
SPI packet EOT
error.
141-SPI POST ERR
SPI packet power on
error.
May be caused by a temporary timing problem in the
communication between the two processors. Power off,
wait one minute, then power on to reset the
communication interface.
if alarm reoccurs, power off the pump and access the test
mode as described in Section 5.3.2.3, Entering the
Service Test Mode. Select 02 CLEAR EXT RAM
then 03 CLEAR ERR LOG. Power off, wait one
minute, then power on again. If alarm does not
reoccur, it has been cleared.
If the alarm reoccurs, check main CPU U19 or safety
CPU U18.
142-CONFIG REG
-
COMFIG register
error which indicates
that the processor is
not properly.
configured.
Return the pump to the manufacturer to have the
processor properly configured or replace the main CPU
U19.
AIM/AIM Pius Technical Service Manual
6-13
430-600035-002 (Rev. 11/99)