Error Messages - Abbott NT000iX Quick Start Manual

Rf generator
Table of Contents

Advertisement

ERROR MESSAGES

ERROR MESSAGE
Probe X over 98 °C, press back*
No thermocouple on probe X*
Check connector and probe
An error has occurred
Emergency stop has been pressed
Fuse blown
Invalid MAC address
Amplifier fault
USB stick not connected
*
"X" indicates the electrode number(s) stated on the generator error message
14
This document is a guide. For complete details, always refer to the NT2000iX
DESCRIPTION
The temperature exceeded 98 °C on one
or more of the electrodes. As a safety
measure, the generator discontinued power
to the electrodes disconnecting the patient
from the machine.
The temperature reading device similar to
a thermostat in the electrode is potentially
damaged or has failed due to wear.
The generator detects a connector but does
not recognize a completed circuit.
The generator was unable to complete the
procedure and disconnected power to the
patient as a safety precaution.
The red "Emergency Stop" button on the
front panel was pressed, disconnecting
power to the electrodes.
An inline fuse in the machine was blown
during the procedure.
The printer's MAC address is invalid
or not set.
A hardware component has
potentially failed.
The generator does not recognize the USB
storage device plugged into the port on the
back of the machine.
RF Generator Operator's Manual.
CORRECTIVE ACTION
Tap Back and then re-enter the lesion
screen and attempt to continue the
procedure. If the problem persists, please
contact your local Sales Representative.
Disconnect the electrode and reconnect.
If the problem persists, exchange the
electrode and contact your local Sales
Representative.
Disconnect the electrode(s) and reconnect.
If the problem persists, exchange the
electrode(s) and contact your local
Sales Representative.
Disconnect all of the electrodes, restart
the machine using the power switch and
attempt to complete the procedure. If the
problem persists, please contact your local
Sales Representative.
Disconnect all of the electrodes, restart
the machine using the power switch and
attempt to complete the procedure. If the
problem persists, please contact your local
Sales Representative.
Please contact your local
Sales Representative.
Ensure that the MAC address of your
printer is entered correctly in the Printer
Set Up screen of your generator. Check
your printer to ensure it is working
correctly and try again.
Please contact your local
Sales Representative.
Remove the storage device, reinsert it
and reattempt the file download. If the
download does not work, the storage
device may need to be formatted. Using
a Windows
based computer, complete a
full format of the storage device. When
complete, reattempt the file download.
If the problem persists, please contact
your local Sales Representative.

Advertisement

Table of Contents
loading

Table of Contents