Section 5, Maintenance and Service Tests
5.3.2.5
Set Clock
TIME
IS 10:42
AM
SAT,
AUG
03,
96
* KEY TO CHANGE
PRESS
NO TO EXIT
_ SET
CLOCK
'YES'=
PASS/NEXT
'NO'
= FAIL/QUIT
or
'BACKUP'
5.3.2.6
Software Version Verification
VERSION
NUMBERS
MAIN
+ tés
SAFETY
Poř
PRESS
NO
TO
EXIT
VERSION
#
\YES'=
PASS/NEXT
'NO'
=
FAIL/QUIT
or
'BACKUP'
5.3.27
Supply Check
SUPPLY
CHECK
EXT
AC
#.##V
###
REG
SV
#.##V
###
PRESS
NO
TO
EXIT
SUPPLY
CHECK
'YES'=
PASS/NEXT
'NO'
=
FAIL/QUIT
or
'BACKUP'
430-600035-002 (Rev. 11/99)
If time and date are correct, press [NO] to exit.
Ifa change is necessary, press the [OPTIONS] key and follow the
instructions on the display to enter the correct information. Press
[NO] to exit when the clock is correct.
If using the PVT Form, record the results.
Press [YES] to indicate PASS and go to the next test, [NO] to exit
the FINAL
ATP
(Acceptance
Test Procedure),
or press
[BACK-UP] to repeat the test.
Record the MAIN and SAFETY software version numbers on the
PVT Form.
Press [YES] to indicate PASS and go to the next test, [NO] to exit
the FINAL ATP, or press [BACK-UP] to repeat the test.
PLEASE WAIT will display while the pump performs the supply
check. PRESS NO TO EXIT appears when the check is complete.
Verify the EXT AC reading is 7.60 - 8.40V,
Verify the REG 5V reading is 4.80 - 5.20V
Press [NO] to exit.
If using the PVT Form, record the results.
Press [YES] to indicate PASS and go to the next test, [NO] to exit
the FINAL ATP, or press [BACK-UP] to repeat the test.
5-10
AIM/AIM Plus Technical Service Manual