Section 5, Maintenance and Service Tests
5.3.4
Optics Test
This test verifies correct operation of the optics by generating an occlusion and an air-in-line alarm.
5.3.4.1
Pump Programming
10.0mL
INFUSED
PRESS
START
TO
INFUSE
REVIEW/RESUME
NEW
CONTAINER
NEW
PROGRAM
CHANGE
PROGRAM
1
TEN
2
PAIN
MANAGEMNT
3
INTERMITTENT
USE
+
FOR
MORE
1
INTRAVENOUS
EPIDURAL
3
SUBCUTANEOUS
CONTINUOUS
BOLUS
ONLY
3 CONT.
+
BOLUS
SELECT
mg/mL
SELECT
ug/mL
3
SELECT
mL
SET
RATE
0.0
mL/hr
ENTER
WHEN
DONE
PROGRAM
A
LOADING
DOSE?
YES
OR
NO
SET
BOLUS
DOSE
0.0
mL
ENTER
WHEN
DONE
430-600035-002 (Rev. 11/99)
Press [CHANGE].
Press [3] to select new program.
Press [2] to select pain management therapy.
Press [2] to select epidural mode.
Press [3] to select continuous plus bolus delivery.
Press [3] to select mL as the delivery unit.
Press [2] [5] and [ENTER] to select a 25 mL/hr rate.
Press [NO].
Press [1] and [ENTER] to select a 1 mL bolus dose.
5-24
AIM/AIM Plus Technical Service Manual