Abbott aim Technical & Service Manual page 45

Table of Contents

Advertisement

5.3.2.12
Battery Check
O
Test Setup:
Section 5, Maintenance and Service Tests
-
Turn the pump power selector on the microbox to BATT BLOCK.
-
Turn voltage selector on the microbox to 5V.
-
Verify that the microbox LED is off.
BATTERY
CHECK
ALK
#.##V
###
LITH
#.##V
dir
PRESS
NO
TO
EXIT
BATTERY
CHECK
'YES'=
PASS/NEXT
"NO"
=
FAIL/QUIT
or
"BACKUP"
5.3.2.13
Idle Current
IDLE
CURRENT
## mA
PRESS
NO
TO
EXIT
IDLE
CURRENT
'YES'=
PASS/NEXT
"NO"
FAIL/QUIT
or
'BACKUP'
AIM/AIM Plus Technical Service Manual
PLEASE WAIT will display while the pump performs the battery
check. PRESS NO TO EXIT appears when the check is complete.
Verify the ALK reading is 4.60 - 5.16V.
Verify the LITH reading is 2.79 - 3.50V.
Press [NO] to exit.
If using the PVT Form, record the results.
Press [YES] to indicate PASS and go to the next test, [NO] to exit
the FINAL ATP, or press [BACK-UP] to repeat the test.
Verify the idle current on the DMM is 30 mA or less (1 mV on the
meter = 1 mA current).
Press [NO] to exit.
If using the PVT Form, record the results.
Press [YES] to indicate PASS and go to the next test, [NO] to exit
the FINAL ATP, or press [BACK-UP] to repeat the test.
5-13
430-600035-002 (Rev. 11/99)