Section 5, Maintenance and Service Tests
5.3.2.21
Print Test Results
Q
Test Setup:
-
Connect the printer to the pump (for setup instructions, refer to the System Operating
Manual).
PRINT
ATP
TESTS:
CONNECT
PRINTER
AND
PRESS
START.
STOP
TO
CANCEL.
5.3.2.22
Printer Test
01
VERSION
#
02
CLEAR
EAT
RAM
03
CLEAR
ERR
LOG
USE
+
FOR
MORE
5.3.2.23
Power Loss Alarm
Press [START] to print the test results.
Attach the printout to the PVT.
At the Service Test Mode menu press [1] [7] to enter the PRINTER
TEST.
The pump will send a copy of the character set to the printer. Verify
the printout reads as follows:
PRINT
TEST
START
ABCDEFGHIJKLMNOPORSTUVWXYZabcdefghijklmn
opgrstuvwxyz0123456789.: *,UABCDEFGHIJKL
MNOPORSTUVWXYZabcdefghijklmnopgrstuvwxyz
0123456789.:
*,UABCDEFGHIJKLMNOPORSTUVWX
PRINT
TEST
COMPLETE
Attach the printout to the PVT.
Disconnect the external power supply and battery substitution block. Verify that the power loss alarm
sounds for at least 30 seconds, Hf using the PVT Form, record the results.
430-600035-002 (Rev. 11/99)