Section 5, Maintenance and Service Tests
5.3.6
Burn-in Test
This optional test verifies correct operation of the pump over 24 hours without alarms.
Note: This test is recommended but not required.
5.3.6.1
Test Setup
1.
Install four (4) AA batteries or power supply list 13586.
2.
Install the test cartridge into the pump.
5.3.6.2
Test Procedure
Switch the pump on and wait for the self-test to complete. Because only batteries are in use, the
following display appears:
USING
BATTERIES
Press [ENTER].
ENTER
TO
CONFIRM
CURRENT
THERAPY:
Press [NO] to enter a new therapy.
PAIN MGMT
EPIDRL
USE CURRENT?
YES OR NO
1 TPN
Press [5] to select continuous therapy.
2
PAIN
MANAGEMNT
3
INTERMITTENT
USE
+
FOR
MORE
SET
RATE
Press [4] [0] [0] and [ENTER] to select a 400 mL/hr rate.
400.0
mL/hr
ENTER
WHEN
DONE
CONTAINER
SIZE
Press [6] [4] [0] [0] [ENTER] to select a 6400 mL container size.
6400.0
mL
ENTER
WHEN
DONE
AIR
SENSITIVITY
Press [3] to turn the air alarm off.
1 HIGH
2 LOW
3 AIR ALARM
OFF
430-600035-002 (Rev. 11/99}
5-30
AIM/AIM Plus Technical Service Manual