Section 5, Maintenance and Service Tests
5.3.7
Preparation for Use
After testing is completed, initialize the NVRAM and clear the logs to prepare the pump for use.
Power on the pump to initiate the self-test.
Enter the Service Test Mode as described in Section 5.3.2.3, Entering the Service Test Mode.
01
VERSION
#
02
CLEAR
EXT
RAM
03
CLEAR
ERR
LOG
USE
*
FOR
MORE
Press [0] [2] to select CLEAR EXT RAM.
CLEARING
EXT
RAM
ADDRESS:
k
The pump clears the external RAM chip. As the test proceeds, the
address changes.
PLEASE
WAIT
No user response required; pump automatically advances. If using
the PVT Form, record the result or "fail" if the procedure fails to
complete successfully. Otherwise, proceed to the next step.
01
VERSION
#
02
CLEAR
EXT
RAM
03
CLEAR
ERR
LOG
USE
+
FOR
MORE
Press [0] [3] to select CLEAR ERR LOG.
CLEARING
The pump clears the error logs. As the test proceeds, the address
ERROR
LOGS
changes.
ADDRESS:
#444
PLEASE
WAIT
No user response required; pump automatically advances. If using
the PVT Form, record the results.
Turn off the pump. Wait at least five seconds, then turn the pump
on.
UNIT
SELF-TEST
IN PROGRESS
TIME
IS ##:##
XM
DDD,
MMM DD,
YY
i
TPN
2
PAIN
MANAGEMNT
3
INTERMITTENT
USE
V
FOR
MORE
430-600035-002 (Rev. 11/99)
If the pump
successfully completes the self-test sequence, the
pump is ready for use.
When the pump displays the first programming screen, turn off the
pump and remove the batteries and/or disconnect the power supply
from the pump.
5-32
AIM/AIM Plus Technical Service Manual