Connection Error When Using The Lds - Primes FocusMonitor FM+ Original Instructions Manual

Laser beam diagnostic system
Hide thumbs Also See for FocusMonitor FM+:
Table of Contents

Advertisement

10.2

Connection error when using the LDS

The firewall may block the connection between the device and
the LDS:
X In Windows > Control panel > Firewall, enable the UDP
port 20034.
The IP address of the PC is not within the range of the device.
X
In Windows > Control panel > Network and Sharing Center,
assign an IP address to the PC that is in the same address
range as the PRIMES device (e.g. 192.168.116.xyz).
The IP address of the PRIMES device can be found on the
identification plate.
If several network cards or a USB3-to-Ethernet card are installed in
the PC, the connection between device and LDS may be blocked
by the selection of the wrong network card.
1.
Select the appropriate network card in the Connections >
All window.
The device is displayed in the Connections window
2.
Click on the device.
3.
Click on the Connect to device button.
66
The UDP port should be enabled by a system
administrator.
The IP address should be set by a system admin-
istrator.