9.9.3

Start measurement

DANGER
Serious injury to the eyes or skin due to laser radiation
If you use a continuous plane measurement to align the device in Monitor measurement mode,
note the following:
X Align the device preferably with a pilot laser that cannot cause dangerous reflections.
X Aligning the device with a class 4 laser can result in dangerous reflections. In this case, the alignment
must be remotely controlled behind a separating protective device. The protective device must block
the radiation or reduce it to a non-hazardous level.
1. Follow the warning messages in
chapter 9.1 „Warning messages" on
page 27.
2.
Click the Start button.
 The measurement begins.
Optional:
X Click the Stop button to abort the
measurement.
X Click the Stop Rotation button to
stop the rotation of the measuring tip.
During the measurement, the progress
is shown in the following indicators:
Measurement:
While the indicator is rotating, the measure-
ment is performed.
3.
Switch off the laser after the measure-
ment is completed, unless you want
to perform further measurements.
9.9.4

Display measurement results

During the measurement, measurement
data is continuously read out and displayed
in the graphical view.
The determined measurement data are not
saved in the project tree of the Projects tab.
X For a detailed description of the file
management and evaluation of the
measured data, please refer to the
separate operating manual LDS.
58
FocusMonitor FM+
Revision 08 EN - 08/2023