Linescan

In measuring mode Linescan, the laser beam is measured at a defined position on the y-axis over a specific
period of time. The line width, line center and position on the y-axis are freely adjustable.
The measuring tip of the FM+ is moved to a fixed y-position. At this position, it measures the power density
on a single measuring path with every rotation of the measurement tip.
The measurement is carried out over a defined period of time or until it is manually aborted.
9.7.1

Search laser beam automatically

1.
Observe the warning messages in
chapter 9.1 „Warning messages" on
page 27.
2.
Click on the Settings tab.
3.
Enter the position of the measure-
ment on the z-axis in the input field
z-position in mm.
4.
Switch on the laser.
5.
Click on the Find beam button.
 The laser beam is automatically
searched for in the entire measuring
range. The measurement window and
gain are set automatically.
If the search is successful, the laser
beam is displayed in the graphic
representation.
If the beam is not displayed:
X Check for correct alignment of the
device on the x-y plane according to
chapter 7.2.3 „Align the device" on
page 19.
X Check the direction of the beam
entrance.
X Adjust the gain.
X Choose a different z-position.
X Increase the laser power (step by step).
6.
If necessary, adjust the measurement
track manually according to the following.
chapter 9.7.2 „Adjust the measurement
track manually" on page 49 .
7.
Start the measurement according to
chapter 9.7.3 „Start measurement"
on page 50.
48
FocusMonitor FM+
Revision 08 EN - 08/2023