FocusMonitor FM+
9.5

Settings that apply in all measuring modes

9.5.1

Define locked area

The Locked area in the Device control > Advanced menu enables a spatial restriction of the measuring
range in y- and z-direction. The defined locked area is not entered by the horizontal and vertical carriers. A
locked area can be set in the individual measuring modes and is adopted in all measuring modes.
For standard installation
The position z=0 and y=0 defines the
lower left corner of the locked area as
park position.
The z-coordinate of the point P1 cannot
be changed.
1.
After opening the Device Control
menu, click the Advanced button.
2.
Use one of the following options:
X Enter the locked area in the P1, P2,
P3 input fields.
X Press the left mouse button and drag
the red dots.
3.
Press the Save device settings button.
 The locked area is saved and applied
to all measurement modes.
Revision 08 EN - 08/2023
z
z=100 mm
Locked area
Laser
y
z=120 mm
Park position
z=0 mm
37