FocusMonitor FM+
9.4.3

Move vertical carrier for setup

The Adjustment area of the Device control menu offers various options for moving the vertical carrier.
The vertical carrier moves from the zero position (z=0) to the entered position.
See chapter 9.12.1 „Distance of the pinhole to the horizontal carrier" on page 64.
Option
z-Position in mm
Move to axis position
Move to axis center position
Tab. 9.1:
Options in the Adjustment area of the Device control menu
Revision 08 EN - 08/2023
Explanation
X Enter the desired z-position in the input field or use the slider.
X Click this button to move the vertical carrier to the z position specified
in the z-Position in mm field.
Note that this does not have to be the maximum traversing range of
your device. If a locked area of the z-axis was stored in the EEPROM
during a previous use of the device, then this locked area is active.
X Click this button to move the vertical carrier to the center position of
the (unlocked) z-axis traversing range.
Note that this does not have to be the maximum traversing range of your
device. If a locked area of the z-axis was stored in the EEPROM during
a previous use of the device, then this locked area is active. The vertical
carrier then moves to the center position of the unlocked range.