FocusMonitor FM+
Remove the measuring tip
1.
Unplug the power cord to prevent the device from being accidentally switched on again.
2.
Turn the FM+ upside down.
3.
Turn the rotational disc until the measuring tip becomes visible in the housing recess.
4.
Remove the fastening screws of the
retaining plate (Torx T8).
5.
Carefully push the measuring tip and
retaining plate upwards through the
housing bore using a screwdriver
6. Pull the retaining plate up slightly and
then forwards until it loosens and can
be removed easily
7.
Carefully remove the measuring tip.
Revision 08 EN - 08/2023