Replace Detector - Primes FocusMonitor FM+ Original Instructions Manual

Laser beam diagnostic system
Hide thumbs Also See for FocusMonitor FM+:
Table of Contents

Advertisement

FocusMonitor FM+
11.4

Replace detector

Generally, the FM+ is equipped with a DFIG-PS+ or DFC+ detector. Detectors with different sensitivity or
different time behavior can be used for special applications.
More information regarding the variety of detectors to ensure an optimum configuration of the FM+ is provided
in chapter 11.5 „Selection of the measuring tip and the detector" on page 73.
NOTICE
Damage of the detector
Touching the sensor surface will damage the detector. This can negatively affect the measuring results.
X Do not touch the detector with your fingers on the sensor surface.
X Do not place the detector on the sensor surface to avoid scratches.
Disassemble detector
1.
Unplug the power cord to prevent the device from being accidentally switched on again.
2.
Remove the plastic screws (D) from
the detector.
3.
Carefully remove the detector from
the position:
X Do not pull the cables!
4.
First loosen the golden angle plug (A),
then the black plug (B).
5.
Remove the detector.
X Do not touch the detector with your
fingers on the sensor surface.
Do not place the detector on the sensor
X
surface to avoid scratches.
Revision 08 EN - 08/2023
D
Touch here
Touch here
Do not compress
here!
Sensor surface
D
B
A