Install the measuring tip
DANGER
Serious eye or skin injury due to laser radiation/Destruction of the measuring tip
If the laser beam hits the measuring tip in the beam path, it could be destroyed and scattered or
directed reflection of the laser beam (laser class 4) occurs.
After installation, move the measuring tip out of the beam path.
1.
Insert the new measuring tip
The inlet aperture is located on the
convex side.
2.
Use up and down movements to
insert the retaining plate into the
rotational disc at an angle of about
45 degrees, ensuring that the guide
groove is pointing upwards. Then
push it down into the recess
X Check whether the retaining plate is
positioned correctly by lightly pressing
the front edge.
3.
Insert the fastening screws (Torx T8)
and tighten hand-tight.
4.
Move the measuring tip out of the
beam path.
If you have rotated the measuring tip, use the arrow buttons to select the "Measuring Tip" setting in
LDS in the menu Device control > Advanced > Measuring tip.
70
FocusMonitor FM+
Pinhole
Fastening screws
Retaining plate
Measuring tip
Revision 08 EN - 08/2023