Primes FocusMonitor FM+ Original Instructions Manual page 38

Laser beam diagnostic system
Hide thumbs Also See for FocusMonitor FM+:
Table of Contents

Advertisement

For upside down installation with rotated measuring tip
The coordinates given as the z-position when measuring always relate to the device coordinate system,
which rotates when the device or measurement tip is rotated.
See chapter 9.5.6 „Coordinate system for rotated measuring tip" on page 42.
The position z=120 and y=0 defines the
upper left corner of the locked area as
park position.
The z-coordinate of the point P1 cannot
be changed.
1.
After opening the Device Control
menu, click the Advanced button.
2.
Use one of the following options:
X Enter the locked area in the P1, P2,
P3 input fields.
X Press the left mouse button and drag
the red dots.
3.
Press the Save device settings button.
 The locked area is saved and applied
to all measurement modes.
Park position necessary for storage of
the device in the transport box:
Since the locked area is stored in the device,
it is no longer possible to approach the park
position at z=120 mm after the device has
been switched off.
X Reset the locked area before turning
off the device.
38
z
z=100 mm
Locked area
Laser
y
FocusMonitor FM+
Park position
z=120 mm
z=0 mm
Revision 08 EN - 08/2023