FocusMonitor FM+
9.6

Device control

9.6.1

Settings

The options listed below in alphabetical order are not available in all measuring modes.
1.
Click on the Settings tab.
2.
Enter the options according to the
explanations in the following table.
X Data with (*) can be saved, see
chapter 9.4.6 „Saving options" on
page 36.
Option
# Pixel in x
autom. Gain
autom. Measurement window
Caustic proposal (z-range in mm)
Deactivate z-axis
Find beam
Revision 08 EN - 08/2023
Explanation
The number of pixels determines the resolution of the measurement.
Use one of the following options:
X Enter a value in the input field.
X Use the slider below the input field.
When using a CO
detector, this option is not available and will not be
2
displayed.
If this option is enabled, the gain is set automatically.
X Set the check mark to enable the option.
If this option is enabled, then the measurement window size will be set
automatically.
X Set the check mark to enable the option.
After a completed caustic measurement, a z-range is displayed in these
fields, whose limits z1/z2 have a distance of ± 3 Rayleigh lengths to the
determined beam focus. This z-range is optimized according to Primes
quality criteria.
X Click on the Apply button to accept the caustic proposal for further
measurements.
With this option the z-axis of the device is deactivated.
This setting is used when the laser system moves to the position of the
next measurement. Enabling the option activates the Manual z-position
in mm input field.
X Set the check mark to enable the option.
This option enables an automatic beam search with an automated mea-
suring window size and measurement. The laser beam is automatically
searched for in the entire measuring range. Measurement window size
and gain are set automatically.
The measured plane is then displayed in the graphic display.
X Click on the button to start the beam search.
Note that the determined measurement data will not be saved in the
project tree of the Projects tab.
43