Installing The Optional Fixed Barcode Reader - Agilent Technologies 1100 Series Reference Manual

Sample capacity extension
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

Installation
2

Installing the Optional Fixed Barcode Reader

The fixed barcode reader can be installed on either the side of the Automation
Interface (the portrait position), or in one of three positions at the rear (the
landscape position). Its position, and how you place the plates in the racks,
will be determined by where you place the barcodes on the plates (see
Figure 25
and
Figure
26).
Barcode reader in portrait position
Barcode
Plate location A1
Rack
Well-plate
Automation
Extension
Handler
Interface
Figure 25
Barcodes on the ends of the plates
The default position is the portrait position. With this arrangement, you can read 60-mm
NO TE
long barcodes with the hand-held barcode reader when the plates are mounted in the rack.
If you want to use the landscape position, and also read the plates with the hand-held
barcode reader when they are in the rack, the barcodes must be no longer than 40 mm.
1100 Series Sample Capacity Extension
27

Advertisement

Table of Contents
loading

This manual is also suitable for:

G2257a

Table of Contents