Clean The Barcode Scanner - Agilent Technologies Magnis G9710A User Manual

Ngs prep system, for in vitro diagnostic use, for research use only, not for use in diagnostic procedures
Table of Contents

Advertisement

Figure 35 Exterior surfaces on the back of the instrument to be wiped down
e Instrument Exterior - Top of the instrument: Using a laboratory wipe moistened with 70%
3 Allow isopropyl alcohol to fully evaporate.
4 From the Touchscreen, initiate a 30-minute UV decontamination procedure. See
decontamination "quick

Clean the barcode scanner

Agilent recommends avoiding all contact with the window of the barcode scanner. You can,
however, clean the window using the instructions below if the window becomes visibly dirty or if
the barcode scanner is not operating well.
CAU T I ON
1 Turn off the instrument at both the power button on the front and the power switch on the
back, and disconnect the power cord from the power supply.
2 Use a soft cloth or wipe moistened with water or a mild detergent-water solution to clean the
window of the barcode scanner. If you use a detergent-water solution, follow with cleaning
using a soft cloth moistened with water or 70% isopropyl alcohol.
Do not touch the window with anything other than the soft cloth or wipe that you are using for
cleaning.
3 Remove any remaining moisture with a dry, soft cloth or wipe.
4 Reconnect the instrument to power supply and turn on the power switch at the back of the
instrument.
Magnis and MagnisDx NGS Prep System User Guide
isopropyl alcohol, wipe the top surface of the instrument
cycle"" on page 42 for instructions.
Review
"Precautions to take when cleaning the system components"
page 52
before proceeding.
"Run a
on
55