Mounting The Clamping Top Onto The Segment Arc; The Setting Of The Clamping System Of The Clamp Mount On The Base Plate - Leica HistoCore AUTOCUT Instructions For Use Manual

Rotary microtome
Table of Contents

Advertisement

7
Troubleshooting

7.4.2 Mounting the clamping top onto the segment arc

2
3
4
Fig. 139
To ensure a proper sectioning result, the clamping top
the segment arc
The clamping is carried out using an eccentric lever
with the setscrew
carried out so that the clamping lever can be rotated to the stop with constantly increasing resistance.
Adjust the clamping with a 2.5 mm Allen key at the setscrew
"slips" when the lever is activated. Continue turning the setscrew
turn to the left or right) then check that the lever no longer "slips", but also does not jam "heavily".
7.5

The setting of the clamping system of the clamp mount on the base plate

1
3
Fig. 140
To set the clamping distance from the eccentric bolt to 270 °, do the following steps.
1. Place the knife holder base
2. Adjust by screwing or unscrewing the countersunk screw
piece in the base using the Allen key No. 3
140-2")
can be clamped at 0 ° position
108
(→ "Fig.
139-4").
(→ "Fig. 139-1")
on the underside of the segment arc. The setting of the clamping is
2
4
(→ "Fig. 140-1")
(→ "Fig. 139-2")
(→ "Fig.
139-3"). The clamping force is adjusted
(→ "Fig. 139-1")
(→ "Fig. 139-1")
5
on the instrument base.
(→ "Fig. 140-5")
(→ "Fig.
140-6"), so that the eccentric bolt
(→ "Fig. 140-3")
and 270 ° position
must be securely clamped onto
so that the lever initially
a little bit (approx. 1/4
6
in the clamping
(→ "Fig.
(→ "Fig.
140-4").