B.5.6

Plumbing the Amperometry Cell

Install the Dionex ED50A amperometry cell directly after the column.
Figure B-15 shows the plumbing schematic for amperometry detection.
A second detector can be installed before the amperometry cell. However,
if a second detector is installed after the cell, make sure the pressure at the
amperometry cell outlet remains below 700 kPa (100 psi). Because of the
volume within the pH reference electrode section of the cell, you may
note some band broadening at the second detector. However, this is
minimized by the precision flat bottomed pH reference electrode.
From Autosampler
Inject Port
To Autosampler
Drip Tray
Figure B-15. Amperometry Cell Plumbing Schematic
Doc. 065551-01 6/13
Cell
Outlet
Amperometry
Union
Analytical
Column
Injection
Valve
4
5
6
3
1 2
From
Pump
Sample Loop
To Waste Container
Waste Line
To Detector's
Connector
Cell Cable
Cell
Cell
Inlet
NOTES:
Route the cell cable and
waste line through the
service chase on the right
side of the compartment.
Guard
Column
B • Installation
B-13