Dionex ED50A Electrochemical Detector
C.1.7
Main Screen—DC Amperometry
The
primary functions in large characters that make viewing easier from a
distance.
DC AMPEROMETRY
- 6. 13
LOCAL
Help Message
Screen Field
Detector
Operating Mode
POTENTIAL
CELL
RANGE
METHOD
Control Mode
MIN
C-10
screen displays the measured current (Amperes) and other
MAIN
nA
METHOD 05
Figure C-9. Main Screen—DC Amperometry
Description
Selects the detector operating mode: Conductivity, DC
Amperometry, Integrated Amperometry, or Voltammetry.
The field on the left sets the polarity. The field on the right sets the
applied potential.
Turns the amperometry cell off or on, disabling or enabling
detection.
Sets the detector sensitivity between 50 pA to 300 A.
Sets the method number. Pressing Select and Enter sets the
detector to Direct control.
Sets the detector to Local, Remote, or Locked Remote control.
Shows the method clock elapsed time.
POTENTIAL
-
CELL
ON
RANGE
1000 nA
34.56 MIN
0.05 V
Doc. 065551-01 6/13