Dionex ED50A Electrochemical Detector
F.1.1

Signal Electrical Parameters

Pin 1
Pin 2
Pin 3
Pin 4
Pin 5
F-2
Attach the negative input of the chart recorder to pin 1.
Select the full-scale output range setting (0 to 0.01 V, 0 to 0.1 V, or
0 to 1.0 V) on the
ANALOG OUT SETUP
Section C.1.15).
The output resistance is 1 to 2 k.
The voltage is equal to that developed across the suppressor, where
the measured voltage equals the suppressor voltage divided by 15.
The suppressor is driven by a regulated current source.
The output resistance is 10 k.
The voltage is proportional to the Dionex DS3 heater current. The
maximum heater power produces approximately 1.2 V.
The output resistance is 5 k.
The voltage is related to the inverse exponential of the temperature.
Table F-2 lists the amperometry flow stream temperatures.
The output resistance is 1 k.
screen (see