B.1 Facility Requirements
Make sure the Dionex ED50A installation site meets the electrical and
environmental specifications listed in Appendix A.
Install the Dionex ED50A on a sturdy table or workbench, at a height that
ensures convenient viewing of the front panel display.
Allow at least 6 cm (2.4 in) clearance behind the Dionex ED50A for power
connections and ventilation.
During installation, or whenever it is necessary to move the detector,
lift it only from the bottom or side. Lifting the Dionex ED50A by the
front panel door will damage the door hinges.
B.2 Power Connection
The power to the Dionex ED50A is controlled from the main power switch on the
Dionex ED50A rear panel. The power supply is auto-sensing; thus, no adjustment
is required to select the line voltage.
Connect the modular power cord (IEC 320 C13), from the main power receptacle
on the rear panel (see Figure B-1) to a grounded, single-phase, power source.
The power supply cord is used as the main disconnect device. Ensure
that the socket-outlet is located/installed near the module and is
easily accessible.
Le cordon d'alimentation principal est utilisé comme dispositif
principal de débranchement. Veillez à ce que la prise de base soit
située/installée près du module et facilement accessible.
Das Netzkabel ist das wichtigste Mittel zur Stromunterbrechung.
Stellen Sie sicher, daß sich die Steckdose nahe am Gerät befindet und
leicht zugänglich ist.
Doc. 065551-01 6/13