Dionex ED50A Electrochemical Detector
C.1.5
Waveform—Integrated Amperometry
Figure C-7 shows the
These entries form points on a plot of potential vs. time.
When the last step displayed is not the last step in the waveform, a lower
case
INT AMP WAVE
STEP
0
1
2
3
4
Help Message
Figure C-7. Waveform Screen—Integrated Amperometry
Screen Field
WAVE EDIT
SAVE TO
RUN
TIME
POTENTIAL
INTEGRATE
C-8
WAVEFORM
is displayed beside the time digits in the last line.
v
EDIT
TIME (sec)
0.00
0.30
0.50
0.51
0.60
V
Description
Specifies the waveform number (0 through 19) to edit.
Specifies the waveform number (0 through 19) to save the
current waveform definition to.
Specifies the waveform number (0 through 19) to run.
Specifies the time for this step. Times are displayed
chronologically.
The left field sets the polarity; the right field is the point in the
waveform for this voltage. If there is no entry, the values from
the preceding step remain in effect.
Sets the begin and ending times for the integration period.
Integration cannot begin on the first waveform step, or end on
the last step.
screen for Integrated Amperometry.
19
SAVE TO
19
POTEN (V)
INTEGRATE
+
0.10
+
0.60
-
0.30
RUN
13
BEGIN
END
Doc. 065551-01 6/13