Dionex ED50A Electrochemical Detector
(DC Amperometry and Integrated Amperometry modes)—Regular
baseline oscillation on high-sensitivity ranges
Reconnect the short length of titanium tubing to the cell inlet.
(DC Amperometry and Integrated Amperometry modes)—Dirty or
pitted working electrode
If a disposable working electrode is being used, replace it. If the working
electrode is non-disposable, polish it (see Section 5.2).
(Integrated Amperometry mode)—Regular baseline oscillations
The water used to prepare the eluent contains trace contaminants. Remake the
eluent, using high purity deionized water containing no contaminants.
4.5

Tailing Peaks

(DC Amperometry and Integrated Amperometry modes)—Dirty or
pitted working electrode
Clean the working electrode with solvent (methanol) and rinse with deionized
water. Dry with a clean soft cloth or tissue.
If a disposable working electrode is being used, replace it. If the working
electrode is non-disposable, polish it (see Section 5.2).
4.6

Amperometry Cell pH Readout Always 7.0

The pH reading is displayed on the
Short circuit in pH reference electrode
Check connections to the cell pre-amp board:
1. Disconnect the inlet and outlet lines from the amperometry cell and
2. Loosen the cell cover thumbscrew and remove the cover.
4-4
disconnect the cell's electrical cable. Remove the cell from its
mounting location in the chromatography compartment.