Dionex ED50A Electrochemical Detector
The first waveform step at 0.00 s is the initial potential of -0.80 V. To
calculate the time for the second step, divide the total voltage scanned by
the desired sweep rate:
The time for the second step is 14 s. The third step is an equal interval
later, or 28 s.
3.7.3

Running the Waveform

When the cell is turned on, the Dionex ED50A applies the initial potential
programmed at time zero. At that point, you may do any of the following:
3-14
Figure 3-7. Cyclic Voltammetry Example
----------------------------------------------------- - = 1.4 V
Total Voltage Scanned
Sweep Rate
To begin the waveform, press
To freeze the scan at the current potential, press
continue from that point.
To return to the initial potential without stopping the scan, press
.
Reset
To return to the initial potential and hold at that potential, press
and
, in that order.
Reset
--------------- - = 14 s
0.1 V/s
.
Run
. Press
to
Hold
Run
Hold