Physical; Environmental; Electrical; Display And Keypad - Thermo Scientific Dionex ED50A Operator's Manual

Electrochemical detector
Table of Contents

Advertisement

A.1 Physical
Dimensions
Weight
Decibel Level
A.2 Environmental
Operating
Temperature
Operating
Humidity
A.3 Electrical
Main Power
Analog Output
A.4 Display and Keypad
Display
Keypad
Doc. 065551-01 6/13
22.5 cm W x 17.0 cm H x 42.0 cm D
(8.8 in W x 6.6 in H x 16.4 in D)
6 cm (2.4 in) clearance required behind the detector
8.2 kg (18 lb)
50 db ("A WEIGHING" setting)
4 ºC to 40 ºC (40 ºF to 104 ºF)
5 to 95% relative humidity, noncondensing
85 to 270 Vac, 47/63 Hz; 40 W Max, 25 W typical. The Dionex
ED50A power supply is auto-sensing and requires no voltage
adjustment.
Fuses
Two 3.15 amp fast-blow IEC127 fuses (P/N 954745)
User-selectable full-scale output of 10, 100, or 1000 mV
Liquid crystal display with adjustable backlighting
26-button keypad for entering commands and numerical values
for screen parameters