Dionex ED50A Electrochemical Detector
C.2.4

Analog Status

This screen reports the status of several analog test points. If a run is in
progress, the status updates are from the last idle state. If no run is in
progress, the values are in real time.
AES/
AES/
AES/
Help Message
Screen Field
AES/SRS
CONNECTED
AES/SRS OVER
TEMP
AES/SRS OVER
VOLT
DS3 CONNECTED
DS3 OVER TEMP
DS3 UNDER TEMP
CD CELL
CONNECTED
EC CELL
CONNECTED
THERMISTOR
TEMPERATURE
C-24
ANALOG STATUS
S
RS CONNECTED:
SRS OVER CURR:
SRS OVER VOLT:
THERMISTOR TEMPERATURE:
Figure C-22. Analog Status Screen
Description
Indicates whether a suppressor (Dionex SRS or AES) is connected
to the detector.
Indicates whether the suppressor is over the temperature specified.
Indicates whether the suppressor is over the voltage range specified.
Indicates whether a Dionex DS3 Detection Stabilizer is connected
to the detector.
Indicates whether the Dionex DS3 is over the temperature specified.
Indicates whether the Dionex DS3 is under the temperature
specified.
Indicates whether a conductivity cell is connected to the detector.
Indicates whether an amperometry cell is connected to the detector.
Indicates the temperature recorded from the conductivity cell
thermistor.
Y
DS3 CONNECTED:
N
DS3 OVER TEMP:
N
DS3 UNDER TEMP:
CD CELL CONNECTED:
EC CELL CONNECTED: Y
nn.n Degrees
Y
N
N
N