Discolored Ph Reference Electrode; Leaking Ph Reference Electrode - Thermo Scientific Dionex ED50A Operator's Manual

Electrochemical detector
Table of Contents

Advertisement

Dionex ED50A Electrochemical Detector
Uncalibrated reference electrode
Calibrate the reference electrode from the
Section 5.4).
Contaminated reference electrode
Soak the reference electrode in a solution containing 1 M KCl and 1 M HCl.
Replace the electrode (P/N 046333) (see Section B.5 for instructions).
Dry reference electrode
Replace the electrode (P/N 046333) (see Section B.5 for instructions).
To prevent a reference electrode from drying out, make sure that mobile phase
is always being pumped through the cell. If the cell will not be used for a short
time (less than 2 days), disconnect the tubing from the inlet and outlet fittings
and install fitting plugs. For longer shutdowns, remove the electrode from the
cell and store it in its storage bottle filled with saturated KCl solution. See
Section 3.9 for detailed instructions.

4.10 Discolored pH Reference Electrode

A slightly discolored reference electrode may continue to function within normal
specifications. However, discoloration indicates reference electrode drift. If you
observe problems with amperometric detection and the reference electrode
appears discolored, perform the potential shift test described in Section 4.12. The
reference electrode typically lasts 6 months in normal use.

4.11 Leaking pH Reference Electrode

Defective pH reference electrode O-ring
Replace the pH reference electrode O-ring (see Section 5.3).
4-6
screen (see
pH CALIBRATION
Doc. 065551-01 6/13

Advertisement

Table of Contents
loading

Table of Contents