consists of a detector interface card, installed in the Dionex ED50A; a computer
interface card, installed in the host computer; and cabling between the two cards.
The DX-LAN interface uses a 10BASE-T RJ-45 (telephone-style) connector. To
install the 10BASE-T cable, see Section B.4.
NOTE For installation instructions for the host computer's
Chromeleon 6.8 runs under the Microsoft Windows XP Service Pack 3 operating
system. When you purchase Chromeleon 6.8 and a PC from Thermo Fisher
Scientific, the chromatography software and computer hardware are installed at
the factory.
B.4 DX-LAN Interface Connections (Optional)
Required Parts
Part
10BASE-T DX-
LAN cable
10BASE-T "combo"
Ethernet hub
The 10BASE-T DX-LAN cable is a Category 5 unshielded twisted-pair
cable. Do not substitute a cable of an inferior grade. Failure to use the
correct cable will cause the detector to lose communication with the
host computer.
When using 10BASE-T cabling, you must use a hub. If you simply
plug the 10BASE-T cable from the detector into the connector on the
host computer's DX-LAN card, the connection either will not work or
will be unreliable.
Doc. 065551-01 6/13
internal DX-LAN card, refer to the Chromeleon 6.8
Help or Installing the Chromeleon Chromatography
Management System with a Dionex Ion Chromatograph
(IC) (Document No. 031883).
Part Number
960281–6 ft
(1.8 m)
057398
Description
Shipped with the detector
Cables in these lengths are also
available: 960279–10 ft (3 m)
960286–1 ft (0.3 m)
Shipped with Chromeleon 6.8