Dionex ED50A Electrochemical Detector
8. Rinse and dry the reference electrode cavity to remove any particulate
9. Verify that the stop ring (P/N 045967) is in place at the bottom of the
Figure 5-5. Reference Electrode Cavity Stop Ring
10. To avoid any hydraulic pressure buildup when inserting the reference
5-6
Figure 5-4. Identifying the CHEMRAZ O-Ring
matter such as salt crystals.
reference electrode cavity (see Figure 5-5). The stop ring prevents the
reference electrode from bottoming out and thereby blocking the flow
path inside the cell.
electrode, make sure that fitting plugs are not installed on the cell
inlet and outlet fittings.