Faulty Dx-Lan Communication - Thermo Scientific Dionex ED50A Operator's Manual

Electrochemical detector
Table of Contents

Advertisement

Dionex ED50A Electrochemical Detector
5. Connect the reference electrode being tested to J2 by turning the
6. Connect the new reference electrode to J2 by turning the electrode at
7. Immerse both reference electrodes in a saturated solution of KCl.
8. When the pH reading has stabilized, calculate the difference in
For most applications, a difference of less than 50 mV is satisfactory.
Therefore, the electrode can still be considered acceptable if the voltage
difference is less than 50 mV. If the difference is greater than 50 mV, try
regenerating the electrode by soaking it in a solution of 1M KCl plus 1M
HCl. If this does not reduce the potential shift, replace the electrode.

4.13 Faulty DX-LAN Communication

DX-LAN interface incorrectly installed
See Section B.3 for 10BASE-T connection instructions.
4-8
electrode at right angles and plugging the center connector of the
three-connector socket into the pin labeled "pH."
right angles and plugging the center connector of the three-connector
socket into the pin labeled "Ag."
potential between the two reference electrodes, using the following
equation:
Difference (Volt) = (7 pH reading) 0.06
Doc. 065551-01 6/13

Advertisement

Table of Contents
loading

Table of Contents