Dionex ED50A Electrochemical Detector
4. Pull the J2 connector through the opening in the electrode cylinder
5. Slide the PEEK O-ring retainer and CHEMRAZ O-ring onto the
6.
While installing the electrode, maintain all parts in a vertical
orientation, with the bottom of the electrode pointing down. This
avoids bubble formation and helps ensure correct centering of the
electrode inside the narrow area at the bottom of the cavity.
B-10
(see Figure B-10).
J2 Connector
Figure B-10. Installing the Electrode Cylinder
electrode (see Figure B-11).
Figure B-11. Installing the O-Ring
Before installing the electrode, observe the inside of the reference
electrode cavity. When installed correctly, the electrode fits into the
narrower diameter area at the bottom of the cavity and sits on top of
the stop ring (see Figure B-7).
O-Ring
Retainer
O-Ring
Doc. 065551-01 6/13