Cleaning the Resistivity
REMOVE TOP COVER
Figure 9-1. Disconnecting the Resistivity Cell
Thermo Scientific
Refer to Figure 9-1 below for this procedure.
Cell
1. Turn off the EASYpure RODI and disconnect it from the power
supply. Remove the power cord.
2. Depressurize the system by opening the unit dispenser draw-off valve,
allowing water to drain until no more flows from the valve.
3. Remove the Ventgard cap.
4. Remove the screws securing the EASYpure RODI top cover.
Warning Depressurize system prior to opening cartridge access door.
Note Ensure you have the correct replacement o-ring available prior to
cleaning resistivity cell.
AND SCREWS
s
CELL
BUSHING
O-RING
MEMBRANE KEY LEADS
Section 9
Maintenance and Servicing
s
AD FROM HERE
DISCONNECT CELL
LEAD FROM HERE
REMOVE FRONT COVER
AND SCREWS
EASYpure RoDi
9-3