2. A recessed lock is located on each side of the fuse holder (see Figure 5-10).
Using a small screwdriver, push each lock toward the center to release it. The
fuse holder pops out slightly when the locks release. When both locks are
released, pull the fuse holder straight out of its compartment.
Locking Spring
Insert screwdriver
and twist to release
3. The holder contains two fuses. Replace both with new 3.15 amp fast-blow
IEC127 fuses (P/N 954745). Thermo Fisher Scientific recommends replacing
both fuses even though only one has failed.
For continued protection against risk of fire or shock, always replace
with fuses of the same type and rating specified in this manual.
Pour maintenir la protection contre les risques d'incendie ou
d'électrocution, remplacez toujours les fusibles par des fusibles du
même type et du même calibre.
Zum
Sicherungswechsel immer Sicherungen des gleichen Typs und mit
gleicher Leistung verwendet werden.
4. Reinsert the fuse holder into its compartment. The fuse holder is keyed to fit
only in its proper orientation. Apply sufficient pressure evenly against the
holder to engage the two locks. The holder is flush against the panel when
both locks are engaged.
5. Reconnect the main power cord and turn on the power.
Doc. 065551-01 6/13
Fuse Holder
(each side)
Main Power
Receptacle
Figure 5-10. Main Power Fuse Holder
Schutz
vor
Feuer
Fuses (2)
Key
Fuse Holder
(side view)
und
Stromschlägen
5 • Service
müssen
beim
5-11