Perform the Self-Tests

Once you have connected all the hardware components for the U4164A module and created a
logical module for U4164A in the Keysight Logic Analyzer application, you are ready to run the
self-tests on U4164A.
1
2
3
4
5
6
7
Keysight U4164A Logic Analyzer Service Guide
Before performing the self- tests, disconnect all probes from the logic analyzer.
Select Help->Self-Test... from the main menu. The Analysis System Self Tests window appears.
From the Select suites list, select the option displayed for U4164A. Then, select All from the Select
tests list.
Select Start to start a complete module self-test. The progress of the self tests is displayed in the
Results area of the window.
When the self-tests are complete, check the Results window to ensure that the Result Summary
says that all tests passed. If all tests did not pass, refer to
Select the Close button to close the Analysis System Self Tests window.
If all self- tests pass, then record "PASS" in the "Logic Analysis System Self-Tests" section of the
Performance Test Record
(page
"To use the flowcharts"
48).
Testing U4164A Performance
on page 54.