2.41 After all three calibration points have been
performed successfully and the detected
temperatures kept the temperatures can be
saved in memory as the new calibration data.
Press any key except EXIT to save the data.
2.42 The "PERFORM CALIBRATION?"
(AUTOCALTC) display will appear. The
calibration date and the temperature detected
for each calibration point will be displayed.
Page 12
2.42.1 If the thermocouple was not properly
removed and is still inserted (stuck) in
solid Tin or Lead, then press F1 to turn
the heater on. When the Tin or Lead
melts, raise the thermocouple to
remove it from the liquid metal. The
heater will turn off automatically and
the "PERFORM CALIBRATION?"
screen will reappear. Clean the
thermocouple tip.
2.43 Press EXIT to return to the MAIN MENU, or
press ENTER to perform another thermo-
couple calibration (begin at Step 2.5).
JFTOT II Calibration Manual Rev. C