Technical Specifications - PAC Alcor JFTOT II Manual

Jet fuel thermal oxidation tester
Table of Contents

Advertisement

2. TECHNICAL SPECIFICATIONS

Setpoint range: 180°C to 380°C
Flowrate range: 1 to 10 mL/min
Flowrate accuracy: ±0.1 mL/min for 2 to 10 mL/min
Thermocouple calibration standards:
0°C ice bath
232°C freezing point of pure Tin (Sn)
327.5°C freezing point of pure Lead (Pb)
Thermocouple type: J (Iron/Constantan)
Thermocouple temperature range: -5°C to +440°C
∆P range: -30 to +400 mmHg (automatically by-
passed at +250 mmHg)
∆P tare timer: 2 min. after heater tube temperature
stable
Operating system pressure range: +450 to +550 psig
Machine I.D.: 1 to 220
Automatic test numbering: 1 to 30,000
Test time range: 4 min. to (500 mL / flowrate) for
example: maximum time at 3 mL/min is 166 min.
Fuel aeration timer: 6 minutes
JFTOT II Maintenance Manual Rev. C
Aeration Flowrate: 1.5 L/Min (Green)
Coolant Flowrate: 10 Gal/Hr (Green)
Heater tube temperature profile timer: 30 min. after
heater tube temperature is stable
Printer interface: Serial RS232, 19200, 9600, 4800 or
2400 Baud, no parity, 1 start bit, 1 stop bit, 8 data
bits DTR handshake flow control uses 25-pin
serial cable between JFTOT and printer trans-
mits standard ASCII format
JFTOT II Temperature range: +5°C to +45°C
Relative Humidity: 0% to 90% non-condensing
Operating power voltage: 115VAC or 230VAC
+15% -10%, single phase
Operating power frequency: 50Hz or 60Hz ±5%
Typical Idle power requirements: 22VA
Typical Operating power requirements: 250VA (at
260°C, 3mL/min, 500psig)
Absolute Maximum power requirements: 700VA
Printer Power Receptacle (Outlet) Rating: 120 VAC,
1.5 Amps.