15. From the qPro Operator main window, click the System Status indicator.
Figure 42. System Status indicator in qPro Operator software
The status window is displayed.
Figure 43. System Status showing Source readings
If the Source VDC or Source Current (or both) show a "fail" status,
Support.
16. Align the instrument optics.
Make sure the system has been on for at least 15 minutes (one to six hours for best
results), and then
align the instrument
Status indicator
optics.
3
Maintaining the Instrument
contact Technical
Nicolet iS5 LS User Guide
49