Thermo Scientific Nicolet iS5 LS User Manual page 21

Table of Contents

Advertisement

Thermo Scientific
5. When the workflow asks you to load hexadecane, lift the sampling arm and use proper
technique to add hexadecane to the sample cup (see "Tips" below).
Figure 14. Nicolet iS5 LS analyzer with sampling arm raised
Tip
• Fill the sample cup with hexadecane (typically 2 drops). Do not overfill the cup.
• Wait a few seconds for the hexadecane in the sample cup to settle before
proceeding with the workflow.
6. Gently close the sampling arm and click continue.
Do not touch the sampling arm until the measurement is completed.
7. Click continue.
After the pathlength and wavelength accuracy tests are completed, the workflow asks you
to
clean the sample
windows. Follow the instructions in the workflow.
Operating the Instrument
Sampling arm in
raised position
Nicolet iS5 LS User Guide
2
17