Thermo Scientific Nicolet iS5 LS User Manual page 49

Table of Contents

Advertisement

Thermo Scientific
NOTICE Never touch a source element with your bare fingers. Skin oils or other deposits
on the element will shorten its life. Always use clean finger cots, gloves, or laboratory
tissue when handling a source element.
To replace the source
1.
Turn off the instrument power switch
2. Disconnect the power cord and the USB cable from the instrument back panel.
3. Remove any sample material from the sample cup and
When you are finished, lower the sampling arm so it engages firmly with the magnet.
4. Carefully set the instrument back on its back so it rests on the back feet.
Figure 36. Nicolet iS5 LS analyzer tipped back for maintenance
and unplug the power cord.
clean the sample
Gently set
instrument on its back
3
Maintaining the Instrument
windows.
Nicolet iS5 LS User Guide
45