Maintaining the Instrument
32
Nicolet iS5 LS User Guide
Time needed: 15 minutes
Tools needed:
• Desiccant replacement kit (or desiccant canisters that have been regenerated)
• 0.05-inch hex key
• #2 Phillips-head screwdriver
• 7/64-inch ball driver or hex key
WARNING Before you replace the desiccant, turn off the instrument power switch and
unplug the power cord. Remove all flammable materials from the area. Entry of
flammable liquids or gases into the instrument is an explosion hazard. If such entry
occurs, contact us immediately and do not apply power to the instrument until the
condition has been corrected.
To replace the desiccant
1.
Turn off the instrument power switch
power strip.
2. Disconnect the power cord and the USB cable from the instrument back panel.
3. Remove any sample material from the sample cup and
When you are finished, lower the sampling arm so it engages firmly with the magnet.
4. Gently set the instrument on its back so it rests on the back feet.
Figure 19. Nicolet iS5 LS analyzer tipped back for service
and unplug the power cord from the wall outlet or
clean the sample
Gently set
instrument on its back
windows.