Maintaining the Instrument
Replacing the Source
44
Nicolet iS5 LS User Guide
WARNING To avoid a burn hazard, use an insulated cloth or hot pad to handle the heated
desiccant canisters. Always wait until the canisters have cooled to room temperature
before installing them in the instrument.
2. Use an insulated cloth or hot pad to remove the desiccant canisters from the oven.
Immediately place the hot canisters in a dry environment such as a desiccator and allow
them to cool to room temperature. As soon as the canisters reach room temperature, you
must either install them in the instrument or seal them in an airtight container or bag.
Otherwise, they will absorb moisture from the air in the room and quickly become
saturated again.
Related Topics
Maintaining the Instrument
Checking the Humidity Indicators
Replacing the Desiccant
Troubleshooting the Instrument
You can replace the instrument's infrared source yourself. To order a source replacement kit,
contact
us.
WARNING
• Before you replace the source, turn off the instrument power switch, unplug the
power cord and wait at least 10 minutes for the source to cool to room temperature.
• Replace the source only with parts supplied by us.
Time needed: 20 minutes
Tools needed:
• Source replacement kit (includes desiccant replacement kit)
• #1 Phillips-head screwdriver
• #2 Phillips-head screwdriver
• 7/64-inch ball driver or hex key
• 0.05-inch hex key
• Clean finger cots, gloves or laboratory tissue